Transvenous Pacemaker Testing

Setup

and

Testing

4

4-23

F1

F2

F3

F4

eyr042.eps

10. Press

F4

XMIT DATA

to transmit data.

11. Press

F1

RESTART

to restart the test.

12. Press

ESC

to return to previous menu(s).

Amplitude of ECG Signal for a Demand Mode
Pacemaker

This test determines the required amplitude of the simulated ECG signal to be
detected by the demand mode pacemaker.

F1

F2

F3

F4

eyr033.eps

1. From the

SELECT TEST

menu, press

F2

SENSE AMP

.