Introduction and Specifications

Instrument

Specifications

1

1-21

Pacer Load Selection.......................................... 1000

Ω

Range ............................................................. 0.05 mV

peak

(50 V

peak

) to 100 mV

peak-to-

peak

Accuracy......................................................... ±5 % of setting
Resolution (Step Size).................................... 0.05 mV

peak

steps from 0.05 mV

peak

to

0.95 mV

peak

0.50 mV

peak

steps from 1.0 mV

peak

to

49.5 mV

peak

05.0 mV

peak

steps from 50 mV

peak

to

100 mV

peak

Default Setting

Amplitude............................................................ 2.0 mV

peak

Width

Range ................................................................. 0.15 mS to 95.0 mS
Accuracy ............................................................. ±5 % of setting
Selection Count .................................................. 50
Resolution (Step Size)........................................ 0.05 mS steps from 0.15 mS to 0.95

mS
1.0 mS steps from 1.0 mS to 19.0 mS
5.0 mS steps from 20 mS to 95.0 mS

Intended Pacemaker Types

AAI (Atrial Pace and Sense Only)
Compatible Pulse Rates ..................................... 30 to 200 PPM

Ventricular Channel

Physiological Simulation

Selection ............................................................. +R-Wave, -S-Wave, and + T-Wave
Rate .................................................................... 30 to 120 BPM, Delayed from the

ventricular demand pacemaker
pulse by 80 % of the pulse-to-pulse
interval or 400 mS (whichever is
shorter).

Active Outputs ...................................................... Selected ventricular test load (4 mm

Banana Jacks) only.

Waveform Selections

Square (SQU)
Triangle (TRI)
Haversine (HSN)
Sine Square (SSQ)
Asymmetrical Triangle (ISO)
(Fixed Width........................................................ 2 mS rise time / 13 mS fall time)
Default Setting .................................................... Sine Square (SSQ)
Test Load(s)