Fluke Biomedical SigmaPace 1000 User Manual

Page 142

Advertising
background image

SigmaPace 1000

Operators Manual

5-12

The Analyzer begins the demand sensitivity test by measuring the pacemaker
pulse rate; then it places the selected waveform at 50 % of the pacemaker
pulse-to-pulse rate interval and initiates the protocol as stated in the main
product specification document.

The Analyzer response displays:

SENSE=XXX,ZZZ.ZZm,SYYYmV.

XXX is the waveform type, as indicated above.

ZZZ.ZZ is the pulse width.

YYY are the three-digit amplitude numeric reading, either with or without the
floating decimal point.

(Example: SENSE=SSQ,100mS,12.5mV for the sine-square selection @ 12.5
mV amplitude and width of 100 mS.)

SETTEST=NOISE

The Analyzer generates a sine wave at the power line frequency of either 50 or
60 Hz at a prescribed amplitude (Range: 0 to 10.0 mV in 0.50 mV steps) upon
the receipt of this command. Both the frequency and the amplitude of this sine
wave are encrypted in the following serial command structure.

The serial command structure is: NOISE=XX,nn.n where the XX represents
the power line frequency in Hz (50 or 60) and nn.n represents the amplitude in
millivolts. Example: NOISE=60,05.5 generates a sine wave at a frequency of
60 Hz at an amplitude of 5.5 mV.

ECG= ON

ECO=OFF

SETTEST=TREND

The Analyzer runs the prescribed trend test using 02 % to 20 % as the
deviation point from the computed average pulse amplitude, rate, and width.

TREND=uu,nnnn,zz,nnn

uu= % test limit for amplitude.

nnnn= Set point for amplitude (0.1 mA resolution.)

Advertising