Fluke Biomedical SigmaPace 1000 User Manual

Page 137

Advertising
background image

Remote Operation

Transcutaneous Pacemaker Remote Setup and Testing

5

5-7

SETMAKE

SETMAKE=nnnn

The specific protocol required for each manufacturer (or make) is selected
using this command. Each protocol contains a set of measurement algorithms
for rate, width, amplitude, and refractory. SETMAKE= followed by the four-
digit alphanumeric entry assigned to each manufacturer completes this serial
command.

The following table shows the SETMAKE protocol for the indicated
manufacturers (Example: SETMAKE = ZM07).

Table 5-1. SETMAKE Protocols by Manufacturer

Manufacturer (Make)

Selection

Default/Physio-Control/Cardiotronics PC01

Hewlett Packard Codemaster

HP02

Marquette Medical

MQ03

Medical Data Electronics

MD04

Medical Research Laboratories

MR05

Zoll Medical PD1200 – 2000

ZP06

Zoll Medical M-Series

ZM07

RDMAKE

This command returns the NAME and Reference # for the specified make.
(Example: ZM-07 = Zoll Medical M-Series pacemaker)

SETLOAD

SETLOAD=nnnn

This command is used to select the desired test load. The above “nnnn”
represents the actual test load resistance value (in ohms) and is based upon the
available factory preset selections.