Remote Operation

Transcutaneous Pacemaker Remote Setup and Testing

5

5-9

RDAMP

The Analyzer transmits the following information string for amplitude from
the data “frozen” with the above HOLD serial command. The data is returned
in the following format: AMP=XXXmA

RDRATE

The Analyzer transmits the following information string for rate from the data
“frozen” with the above HOLD serial command. The data is returned in the
following format: RATE=XXXPPM

RDWIDTH

The Analyzer transmits the following information string for width from the
data “frozen” with the above HOLD serial command. The data is returned in
the following format: WIDTH=XXXmS

RDENERGY

Even though the pacemaker pulse energy reading is not displayed when the
RDMETER serial command is utilized, it is computed. The Analyzer transmits
information from the data “frozen” with the HOLD serial command.

One of the following formats is used, depending on the energy level of the
transcutaneous pacemaker pulse.

ENERGY= XXX?J

ENERGY=XXXmJ

ENERGY=X.XXJ

RESUME

Releases the HOLD serial command and proceeds with making measurements.

SETTEST=DEMAND

The Analyzer enters the qualitative demand test. The applied pacemaker pulse
rate is measured; if it is ZERO, the Analyzer transmits RATE=000 and will not