Operators Manual

4-34

DC Load Test

F1

F2

F3

F4

eyr034.eps

1. From the

SELECT TEST

menu, press

F2

DC LOAD

.

F1

F2

F3

F4

eyr083.eps

2. Make necessary connections, as follows:

a. Plug the supplied 9 V dc Load Test Cable into the matching

connector on the left side of the Analyzer.

b. A 9 V dc Load Test Cable (Fluke Biomedical Part# 2201074) is

supplied as a standard accessory with the Analyzer. It is designed
specifically for the testing of Medtronic models 5388 Dual Channel
and 5348 Single Channel transvenous pacemakers.