Fluke Biomedical SigmaPace 1000 User Manual

Page 82

Advertising
background image

SigmaPace 1000

Operators Manual

3-28

flexibility allows the user to enter either the actual test pacemaker values or the
selected rate and amplitude settings.

Deviation values are independently selectable from 2 % to 20 % in 1 %
increments. The Analyzer compares these incoming pacer pulses to the target
values. When the long-term test is complete, output pulses from the
pacemaker that were not within the selected limits may be reviewed.

F1

F2

F3

F4

eyr013.eps

1. From the

SELECT TEST

menu, press

F2 LONG TERM

.

F1

F2

F3

F4

eyr026.eps

2. Press

F2 RATE

to toggle between pulse rate and amplitude. This

parameter works with the

LIMIT

parameter to assign limits to both.