QC5200 Series User’s Guide

Counter setup

The counter window will display changing displacements as the axes are moved. Each axis is given a

default encoder resolution. The axis resolutions can be changed to indicate accurate axis displacements

during calibration.

NOTE

This feature is only provided as a convenience to the user, will not be retained when

the program is closed and is not necessary for encoder calibration.

To change the counter display:

1 Click Encoder Resolution

to display the data entry

fields.

2 Enter the axis encoder resolutions into the Axis fields and click OK

to return to the Encoder Setup screen.

Limit Switch setup

Limit switches can be used to prevent axis travel beyond predetermined

points. Axis limit switches can be configured to be active when low

(0 volts), or to be active when high (5 volts).

To configure axis limit switches:

1 Click Limit Switch Setup

to display the limits which

window.

2 Enter 0 for axis limit switches that will be active when low (0 volts).

Enter 1 for axis limit switches that will be active when high (5 volts).

3 Check Swap Limit Switches to exchange the Plus and Negative

switches of an axis.

4 Check the Enable box to enable limit switches for each axis.

5 Click OK to return to the Encoder Setup screen.