Problem, Probable cause, Recommended solution – HEIDENHAIN IK 5294 User Manual
Page 311: Probable cause #1, Recommended solution #1
Advertising
Problems, Probable Causes & Solutions
291
1
1
Problem Solving
Problem
Cannot drag data fields from the Results window into a template
Probable cause
• Templates have been locked to prevent modifications
Recommended solution
• Use the Tools/Options/Locks tab to unlock
the templates
Problem
Changes to the QC5200 settings, window positions, etc. are not saved
Probable cause #1
• The Save on exit option is not enabled on the Tools/Options/General tab
Recommended solution #1
• Use the Tools/Options/General tab to enable
the Save on exit option
Clear check
mark to unlock
templates
Check to enable
Save on exit