Pre-selecting features for a construction, Displaying distances as absolute values, Automatically complete point measurements – HEIDENHAIN IK 5294 User Manual

Page 283

Advertising
background image

263

1

0

Supervisor Setup

Pre-selecting features for a construction

Two methods can be used to construct features:

Method 1 Method 2

• Click a measure tool • Select existing features

• Select existing features • Click a measure tool

• Click measure OK • Click measure OK

The required steps and resulting constructions are the same in either case. However, the order of steps can

be specified to suit the preferences of the user. Check the Allow

pre-selection... box to select features before choosing a measure

tool.

Displaying distances as absolute values

Distance values can be positive or negative depending on the locations of features selected for distance

constructions, and on the order of selection. However, the system can be configured to display the absolute

values (always positive) of distances regardless of feature location and selection order. Check the Always

display distance... box to display distances as positive values.

Automatically complete point measurements

Normally, point measurements are completed by clicking OK in the Feature measure dialog box after

probing the point. This last step can be eliminated for point measurements. Check the Automatically fin-

ish a point measurements... box to automatically complete point

measurements.

Specifying the maximum number of measurement points

Each probe can be configured collect a specific maximum number of points when it’s fired. These points

accumulate as a probe is repeatedly fired during a feature measurement. The system can be configured to

collect up to 3000 points for an entire feature measurement. Enter the maximum number of points for a

complete feature measurement into the Maximum number... data

field.

Specifying the minimum number of measurement points

The feature fitting algorithms require a minimum number of probed points to determine

each feature type’s location and size. The default values shown in the Minimum Points

data fields reflect these minimums. However, the system can be forced to require a

greater minimum number of points to increase the measurement accuracy when the mini-

mum number of points is probed. Enter the desired values into the fields provided.

Measurement Setup

10 Supervisor setup OE.indd 23

5/9/2005 10:04:09 AM

Advertising
This manual is related to the following products: