QC5200 Series User’s Guide

4 Select a new magnification for offset compensation. In this example,

the X20 magnification is selected.

5 Measure the reference

circle again using the new

magnification.

An offset might be shown in

the Part View window.

6 Construct a distance

between the reference circle

and the new circle. Refer

to Chapter 5: Measuring for

instructions regarding con-

structions if necessary.

7 Click Crosshair Properties and enter the X and Y

distance values shown in the Results window into the X

and Y offsets fields for the new probe, and then click

OK.

The position data gathered using the new probe magnification will now match those of the reference probe

magnification in subsequent measurements.

8 Repeat the process of steps 4 through 7 to compensate other probe magnification offsets.