Optical edge detection probes

Probe magnifications can be added and deleted from the Probe menu as optical lenses are installed and

removed.

Edge recognition should be calibrated at the light levels of all optical magnifications. Position offsets

between the crosshair and optical edge probes can be eliminated for all optical magnifications using cross-

calibration functions. Position offsets resulting from minor mechanical lens misalignments can be elimi-

nated using magnification offset compensation functions. CNC probing parameters can be specified to

control probing velocity and data point search distance.

Adding and deleting probe magnifications

Probe magnifications are added

and deleted using the Tools/

Options/Optical Probes setup

screen. Refer to Chapter 10: Su-

pervisor Setup for details.

Calibrating optical edge recognition

Optical edge recognition depends on the contrast between dark and light areas of the part image, and much

less commonly, on the edge characteristics of the part being measured.

Typical optical edge calibrations only involve simple measurements of dark and light areas of the screen at

each optical magnification. However, special part edge characteristics might occasionally necessitate the

measurement of the dark/light contrast boundaries at the part edge. The Teach function performs simple

measurements of dark and light areas of the screen while the Distance Cal function performs contrast

measurements at the part edge.