Agilent Technologies Agilent E5250A User Manual

Page 101

Advertising
background image

Agilent E5250A User’s Guide, Edition 11

5-7

Controlling the E5250A

Switch Control Basics

Table 5-2

Card Number

In the channel_list, you can specify multiple channels by using comma (,) or colon
(:) as follows:

Comma: use between each specified channel as in following examples:

(@10101,10102,10103) means 10101, 10102 and 10103.

(@10112,10202) means 10112 and 10202.

(@11010,20102) means 11010 and 20102.

Colon: use to specify a range of channels as in following examples:

(@10101:10103) means 10101, 10102, and 10103.

(@10112:10202) means 10112, 10201, and 10202.

(@11012:20102) means 11012, 20101, and 20102.

As shown in last two examples, you can specify a range across input ports and card
slots.

NOTE

In the Auto Configuration mode, you can omit any zeros at beginning of channel.
For example, channel 00101 can be expressed by 101.

Channel Configuration

you specified

Slot No. Card is Installed

Card No.

Normal Configuration

1

1

2

2

3

3

4

4

Auto Configuration

1 to 4

0

Advertising