General specifications – Agilent Technologies Agilent E5250A User Manual

Page 297

Advertising
background image

Agilent E5250A User’s Guide, Edition 11

10-3

Specifications

General Specifications

General Specifications

Temperature

Range

Operating: 5

°C to 40 °C

Storage:

−40 °C to 70 °C

Humidity Range

Operating: 5 % to 80 % RH, non-condensing
Storage: 5 % to 90 % RH at 65

°C

Altitude

Operating: 0 to 2,000 m (6,500 ft)
Storage: 0 to 15,240 m (50,000 ft)

Regulatory

Compliance

Safety: CSA C22.2 No. 1010.1 / IEC61010-1

EMC: CISPR 11 Group 1 class A&EN50082-1

Power

Requirement

100 to 240V ± 10 %, 47 to 63 Hz, maximum 100 VA

Number of Slots

4 slots for 30-mm height switch cards

Dimensions

Mainframe: 230 mm H

× 430 mm W × 600 mm D

Option 001: 30 mm H

× 395 mm W × 424 mm D

Option 501: 30 mm H

× 395 mm W × 424 mm D

Weight

Mainframe: 11.2 kg
Option 001: 2.0 kg
Option 501: 2.1 kg

Number of Ports

with option 001

Low Leakage I-V Port: 2
General I-V Port: 4
C-V Port: 2
HF Port: 2
Output Channels: 12

Number of Ports

with option 501

Low Leakage I-V Port: 3
Bias Port: 3
Output Channels: 24 (8

Ч 3)

Advertising