Option 501 multiplexer, Switch specifications – Agilent Technologies Agilent E5250A User Manual

Page 300

Advertising
background image

10-6

Agilent E5250A User’s Guide, Edition 11

Specifications
Option 501 Multiplexer

Option 501 Multiplexer

Switch Specifications

This specification is defined with the module plugged into the mainframe. The value
depends on the following temperature and humidity conditions.

A: 23

°C ± 5 °C, 5 % to 60 % RH

B: 5

°C to 18 °C, 28 °C to 40 °C, 5 % to 60 % RH

C: 28

°C to 40 °C, 60 % to 80 % RH

A B C

Max. Current Rating (A)

1.0

1.0

1.0

Max. Voltage Rating (V)

Channel to Guard

200

200

200

Channel to Channel

300

300

300

Closed Channel Residual Resistance (

Ω)

Low Leakage I-V Port

0.6

0.6

0.6

BIAS Port

a

1.0

1.0

1.0

Channel Isolation Resistance (

Ω)

Low Leakage I-V Port

10

13

5

Ч 10

12

2

Ч 10

12

BIAS Port

10

9

5

Ч 10

8

2

Ч 10

8

a. When 0

Ω resistors are mounted on the card as the protective resis-

tor.

Advertising