To use the VFP data upload library

If you want to use your VFP data in the program, change the program as follows.
Before executing the program, you need to create the VFP data and save it to a file
on diskette. And you must link the VFP Data Upload Library to the program. For
details, refer to “Using the Virtual Front Panel (VFP) Utility” on page 5-4 and
“Using the VFP Data Upload Library” on page 6-22.

1. Insert an exclamation mark (!) at the beginning of each program line shown in

the following table, and enter the lines.

Subprogram Program

Line

Meas_vth
"Constant Definition" block

Test_comm$="MOSFET Vth Measurement"
Channel$="(@10101,10202,10303,10404)"

Meas_vth
"Connect to DUT" block

OUTPUT @Hp5250;":ROUT:CLOS "&Channel$
OUTPUT @Hp5250;"*OPC?"
ENTER @Hp5250;A

Meas_vth
"Connection All Open" block

OUTPUT @Hp5250;":ROUT:OPEN:CARD ALL"
OUTPUT @Hp5250;"*OPC?"
ENTER @Hp5250;A

Meas_c
"Constant Definition" block

Test_comm$="Capacitance Measurement"
Channel$="(@10905,11006)"

Meas_c
"Connect to DUT" block

OUTPUT @Hp5250;":ROUT:CLOS "&Channel$
OUTPUT @Hp5250;"*OPC?"
ENTER @Hp5250;A

Meas_c
"Connection All Open" block

OUTPUT @Hp5250;":ROUT:OPEN:CARD ALL"
OUTPUT @Hp5250;"*OPC?"
ENTER @Hp5250;A

Set5250init

OUTPUT @Hp5250;":ROUT:FUNC NCON"
OUTPUT @Hp5250;":ROUT:BIAS:STAT ALL,OFF"
OUTPUT @Hp5250;":ROUT:CONN:RULE ALL,SROU"
OUTPUT @Hp5250;":ROUT:CONN:SEQ ALL,BBM"
OUTPUT @Hp5250;":ROUT:OPEN:CARD ALL"
OUTPUT @Hp5250;"*OPC?"
ENTER @Hp5250;A