Agilent Technologies Agilent E5250A User Manual

Page 63

Advertising
background image

Agilent E5250A User’s Guide, Edition 11

3-13

Executing Self-Test and Leak Test

Using the Self-Test Utility

Executing the Controller Test

To execute the controller test, press the CONTROLLER softkey.

If [F] result is displayed after the test, the E5250A is defective. Contact your
nearest Agilent Technologies Service Center.

Executing the Front Panel Interface Test

To execute the front panel interface test, do the following:

1. Press the FRONT PANEL softkey.

2. Confirm that the following four LEDs on the E5250A front panel are blinking.

And press the

Local/Self Test

key on the E5250A front panel within 10 seconds.

Remote LED

System Fail LED

Fail LED

Local/Self Test LED

NOTE

Fail Condition for Front Panel Interface Test

If an LED does not blink after selecting the FRONT PANEL softkey, the LED is
defective.

If you do not press the

Local/Self Test

key within 10 seconds, the Fail LED turns on.

If you pressed

Local/Self Test

key within 10 seconds, and the Fail LED still turns on,

it means the

Local/Self Test

key is defective.

Contact your nearest Agilent Technologies Service Center.