[:ROUTe]:BIAS:PORT?

[:ROUTe]:BIAS:PORT?

NOTE

This command applies only to the E5252A card.

For the E5255A card, the input Bias Port cannot be selected, and is always the
built-in Bias Port.

This query command returns the input Bias Port number for the specified card.

Syntax

[:ROUTe]:BIAS:PORT? card_number

Query response

port_number <newline><^END>

port_number = 1 to 10

Example

OUTPUT @Hp5250;":ROUT:BIAS:PORT ALL,4"

OUTPUT @Hp5250;":ROUT:BIAS:PORT? 1"

ENTER @Hp5250;A
For above example, A will be as follows:

4

Parameter Explanation

card_number

card number:

For Auto Config mode: 0

For Normal Config mode: 1, 2, 3, or 4