4-12

Agilent E5250A User’s Guide, Edition 11

Setting up Measurement Environment
Mounting Connectors Directly

NOTE

To Check Interlock Circuit

If you use the 4155/4156, you can easily check the interlock circuit as follows:

1. Connect the Intlk connector of the 4155/4156 to your interlock circuit.

2. Press

System

front-panel key, then select CALIB/DIAG primary softkey to

display the SYSTEM: SELF-CALIBRATION/DIAGNOSTICS page.

3. In the CALIB/DIAG field, select DIAG secondary softkey.

4. In the CATEGORY field, select I/O PERIPH secondary softkey.

5. Move pointer to the 403 (INT.) Interlock LED field.

6. Select EXECUTE secondary softkey.

7. Confirm the following:

LED turns on within 1 sec from when interlock circuit is shorted.

LED turns off within 1 sec from when interlock circuit is open.

To stop the interlock test, select STOP secondary softkey.