Agilent Technologies Agilent E5250A User Manual

Page 116

Advertising
background image

5-22

Agilent E5250A User’s Guide, Edition 11

Controlling the E5250A
Switch Control Basics

In the Auto Configuration mode, you can omit any zeros at beginning of channel.
For example, channel 00101 can be expressed by 101.

Figure 5-7, Table 5-9 and Figure 5-8 show examples of the Channel Number. In this
example, four E5255As are installed in the E5250A. And the E5255As are
configured to make the following multiplexers:

Slot1 and Slot2:

all six blocks are used together to make a 2

×48 multiplexer.

IV input port is internally connected to SMU1 input port.

Slot3 and Slot4:

Both Block1 used together to make 2

×16 multiplexer-1.

IV input port is internally connected to SMU2 input port.

Both Block2 used together to make 2

×16 multiplexer-2.

IV input port is internally connected to SMU3 input port.

Block3 of Slot3 makes 2

×8 multiplexer-1.

IV input port is internally connected to SMU4 input port.

Block3 of Slot4 makes 2

×8 multiplexer-2.

IV input port is internally connected to SMU5 input port.

For the internal connections of the E5255A, see Chapter 2.

Figure 5-7

Example Setup of E5255A

2x48 Multiplexer

slot 2

slot 3

slot 4

slot 1

2x16 Multiplexer-1

2x8 Multiplexer-1

2x8 Multiplexer-2

2x16 Multiplexer-2