Executing Self-Test and Leak Test

Executing Self-Test

Table 3-2

SCPI Commands for Self-Test

NOTE

Front Panel Interface Test

After the :DIAG:TEST:FRAM:EXEC? FPAN command is executed, the Remote,
System Fail, Fail, and Local/Self Test LEDs on the E5250A front panel blink for
about 10 seconds. If an LED does not blink, the LED is defective.

Confirm that the four LEDs are blinking, then press the

Local/Self Test

key on the

E5250A front panel within 10 seconds. The LEDs should stop blinking.

If you do not press the

Local/Self Test

key within 10 seconds, the E5250A returns a

fail result for the Front Panel Interface Test, which turns on the Fail LED.

If you pressed

Local/Self Test

key within 10 seconds, and the Fail LED still turns on,

it means the

Local/Self Test

key is defective.

If you find any problem, contact your nearest Agilent Technologies Service Center.

Test Item

Description

SCPI Command

Controller
Test

Executes Controller Test,
then returns test result.

:DIAG:TEST:FRAM:EXEC? CONT

Returns test result.

:DIAG:TEST:FRAM:STAT? CONT

Clears test result.

:DIAG:TEST:FRAM:CLE CONT

Front Panel
Interface
(I/F) Test

Executes I/F Test, then
returns test result.

:DIAG:TEST:FRAM:EXEC? FPAN

Returns test result.

:DIAG:TEST:FRAM:STAT? FPAN

Clears test result.

:DIAG:TEST:FRAM:CLE FPAN

Relay Test

Executes Relay Test, then
returns test result.

:DIAG:TEST:CARD:EXEC? ALL
or Card No. (1, 2, 3, or 4)

Returns test result.

:DIAG:TEST:CARD:STAT? ALL
or Card No. (1, 2, 3, or 4)

Clears test result.

:DIAG:TEST:CARD:CLE ALL or
Card No. (1, 2, 3, or 4)