Couple port, Couple port -13 – Agilent Technologies Agilent E5250A User Manual

Page 107

Advertising
background image

Agilent E5250A User’s Guide, Edition 11

5-13

Controlling the E5250A

Switch Control Basics

Couple Port

NOTE

This function is available only for the E5252A.

Bias Port and Couple Port cannot be set to the same port number.

This function is used for making Kelvin connections. Couple port allows
synchronized connection of two adjacent input ports to two adjacent output ports.
You set up the Couple Port by using the [:ROUTe]:COUPle:PORT command
described in Chapter 7.

To turn Couple Port mode ON/OFF, refer to [:ROUTe]:COUPle[:STATe] command
in Chapter 7.

For a programming example, refer to Chapter 6.

The input ports 1 and 2, 3 and 4, 5 and 6, 7 and 8, or 9 and 10 can be coupled
together. See Table 5-4. For example, if you specify input port number 1 in the
[:ROUTe]:COUPle:PORT command, input ports 1 and 2 are coupled together.

Table 5-4

Couple Port Number and Output Port Number

Couple

Port No.

Input Ports

Available Output Port No.

01

SMU1, SMU2

01 to 12 (for Normal Config.)
01 to 48 (for Auto Config.)

03

SMU3, SMU4

01 to 12 (for Normal Config.)
01 to 48 (for Auto Config.)

05

SMU5, SMU6

01 to 12 (for Normal Config.)
01 to 48 (for Auto Config.)

07

HF1, HF2

01 to 12 (for Normal Config.)
01 to 48 (for Auto Config.)

09

CV1, CV2

01 to 12 (for Normal Config.)
01 to 48 (for Auto Config.)

Advertising