Hpe5250a_coupleport – Agilent Technologies Agilent E5250A User Manual

Page 244

Advertising
background image

8-20

Agilent E5250A User’s Guide, Edition 11

VXIplug&play Driver
Driver Functions

hpe5250a_couplePort

This function sets the couple ports which are used for making kelvin connections on
the specified card. The specified input port number will be coupled with the next
input port and two output ports. For each card, you may setup the same or different
couple ports. This command overwrites the previous couple port setting for the card.
This command applies only to the E5252A card.

The couple port mode is controlled by the hpe5250a_coupleState function.

Syntax

ViStatus _VI_FUNC hpe5250a_couplePort(ViSession vi,
ViInt16 coupleport_cardno, ViInt16 port1, ViInt16 port3, ViInt16 port5,
ViInt16 port7, ViInt16 port9);

Parameters

vi

Instrument handle returned from hpe5250a_init( ).

coupleport_cardn
o

Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5
(all card) in the normal configuration mode, or 0 (all card in the
automatic configuration mode). For the configuration mode, see
hpe5250a_func.

port1

Couple port by the input ports 1 and 2.
0 (disable) or 1 (enable).

port3

Couple port by the input ports 3 and 4.
0 (disable) or 1 (enable).

port5

Couple port by the input ports 5 and 6.
0 (disable) or 1 (enable).

port7

Couple port by the input ports 7 and 8.
0 (disable) or 1 (enable).

port9

Couple port by the input ports 9 and 10.
0 (disable) or 1 (enable).

Advertising