Figure 2-2 – Agilent Technologies Agilent E5250A User Manual

Page 43

Advertising
background image

Agilent E5250A User’s Guide, Edition 11

2-11

Installation

Installing the E5250A

Figure 2-2

Agilent E5255A Component Locations

To Modify BIAS Input

Each multiplexer block has an independent BIAS input. You can change the BIAS
input internally using the BIASin/BIASout terminals shown in Figure 2-2, or
externally using the BIAS INPUT connectors. For actual examples of modifications,
see “E5255A Configuration Examples”.

To Use Multiplexer Blocks on another E5255A

When a module is installed in a mainframe slot, six module pins are connected to
the mainframe internal bus. So, if multiplexers on different E5255As use the same
module pin, the IV inputs are connected together via the internal bus.

SM

U6

SM

U5

SM

U4

SM

U3

SM

U2

SM

U1

IVout3
connector

IVout2
connector

IVout1
connector

IVin1
cable

IVin3
cable

IVin2
cable

DIP SW2

DIP SW3

DIP SW1

Resistors
(Total 24)

Resistor holder 1

Screws
(Total 48)

Block 1

Block 3

Block 2

Resistor holder 3

Resistor holder 2

BIASin2
terminal

BIASin3
terminal

BIASout1

terminal

BIASout2

terminal

SMU input connectors

BI

T1

(1

)

BI

T3

(4

)

BI

T4

(R

)

BI

T2

(2

)

Resistors

BIT4

0 ohm

0

other

1

To BIAS INPUT 51

To BIAS INPUT 52

To BIAS INPUT 53

IVin cable

DIP SW setting

connection

BIT1 BIT2 BIT3

SMU1

1

0

0

SMU2

0

1

0

SMU3

1

1

0

SMU4

0

0

1

SMU5

1

0

1

SMU6

0

1

1

1: ON

0: OFF

Module pins

Advertising