E5252A Channel List

This section describes how to define the channel_list specifically for the E5252A.

For the basic syntax of the channel_list, see “Channel List and Channel
Configuration” on page 5-6.

Normal Configuration Mode

To specify a channel, you use five digits as shown in Table 5-5.

Table 5-5

Specifying E5252A Channels in Normal Config. Mode

Card Number:

One digit: 1 to 4. E5250A slot number in which card is
installed.

Channel Number: Four digits: XXYY.

XX: input port number (01 to 10).

YY: output port number (01 to 12).

Channel Number assignment is shown in Figure 5-4.

NOTE

For the following two cases, three ports share one input path. Do not specify two of
the input ports at same time.

SMU5, HF1, and CV1 (Input Port No. 05, 07, and 09)

SMU6, HF2, and CV2 (Input Port No. 06, 08, and 10)

For example, if you specify input port 05, then you cannot specify input ports 07 and
09. You can specify 05 and 06 at the same time.

Slot No.

(E5252A is

installed)

One Channel: 5 digits

Card Number

(first 1 digit)

Channel Number: 4 digits

Input Port No.

(next 2 digits)

Output Port No.

(last 2 digits)

1

1

01 to 10

01 to 12

2

2

01 to 10

01 to 12

3

3

01 to 10

01 to 12

4

4

01 to 10

01 to 12