Agilent E5250A User’s Guide, Edition 11

6-31

Programming the E5250A

Using the Capacitance Compensation Routine

1. Measure R, L, C of your cable using the 4284A. Measurement setup is as

follows:

2. Overwrite the R, L, C data in the program lines listed below. The data must be in

per meter values.

Original program defines initial data by DATA statement in the "Rlc_data" block
as shown below:

1350 Rlc_data: !

1360

!

R [ohm]

L [H]

C [F]

:

:

:

:

1420

DATA 100.70E-3, 400.00E-9, 80.00E-12 !User Triax Cbl H

1430

DATA 100.70E-3, 400.00E-9, 80.00E-12 !User Triax Cbl L

1440

DATA 114.00E-3, 544.00E-9, 130.00E-12 !User Coax Cbl H

1450

DATA 114.00E-3, 544.00E-9, 130.00E-12 !User Coax Cbl L

:

:

:

:

FORCE or SENSE (A)

FORCE or SENSE (B)

Insulator

Triaxial Cable

GUARD (C)

GUARD (D)

GROUND (F)

GROUND (E)

Measurement

Parameter

Measurement

Frequency

Measurement

Function

Measurement

Terminals

a

R

1 kHz to 1 MHz

b

A and B

L

SERIES

see note

c

C

PARALLEL

A and C

a. See figure above.
b. Select 1 point in the frequency range.
c. For triaxial cable, connect B to F directly, and measure L

between A and E.
For coaxial cable, connect B to D directly, and measure L
between A and C. Ignore E and F.

Program Line No.

Defines R, L, C value of

1420

Triaxial cable for C meter High terminal

1430

Triaxial cable for C meter Low terminal

1440

Coaxial cable for C meter High terminal

1450

Coaxial cable for C meter Low terminal