Agilent E5250A User’s Guide, Edition 11

6-7

Programming the E5250A

Creating a Control Program

Table 6-1

Programming Summary

For details about these parameters, refer to Chapter 5 and Chapter 7.

CAUTION

Bias Port and Couple Port cannot be set for the same input port number.

Description

Command Default

Setting

1

Defines Channel Configuration Mode

:ROUT:FUNC channel_config

NCON

2

Defines Connection Rule

:ROUT:CONN:RULE card_no, rule

ALL,FREE

3

Defines Connection Sequence

:ROUT:CONN:SEQ card_no,sequence

ALL,BBM

4

Using Bias Mode

Selects input Bias Port (for E5252A)

:ROUT:BIAS:PORT card_no, input_port_no

ALL,10

Bias enables all output ports

:ROUT:BIAS:CHAN:ENAB:CARD card_no

ALL

Bias enables specified output ports

:ROUT:BIAS:CHAN:ENAB:LIST channel_list

All enabled

Bias disables all output ports

:ROUT:BIAS:CHAN:DISA:CARD card_no

All enabled

Bias disables specified output ports

:ROUT:BIAS:CHAN:DISA:LIST channel_list

All enabled

Sets Bias Mode ON/OFF

:ROUT:BIAS:STAT card_no, state

ALL,OFF

5

Using Couple Port (for E5252A)

Sets up Couple Port

:ROUT:COUP:PORT card_no, 'input_port_no'

No couple ports

Sets Couple Port Mode ON/OFF

:ROUT:COUP:STAT card_no, state

ALL,OFF

6

Controls Relays on Cards

:ROUT:CLOS channel_list

All relays open

:ROUT:OPEN channel_list

All relays open

:ROUT:OPEN:CARD card_no

ALL

Parameter Description

channel_config

NCON or ACON.

rule

FREE or SROU.

sequence

NSEQ, BBM or MBBR.

state

ON (1) or OFF (0).

channel_list

5 digit decimal number. See Chapter 5.

card_no

ALL or 0 or 1 to 4 are available.

input_port_no

1 to 10 are available for :ROUT:BIAS:PORT.
1, 3, 5, 7 or 9 are available for :ROUT:COUP:PORT.