7-18

Agilent E5250A User’s Guide, Edition 11

Command Reference
:DIAGnostic:TEST:FRAMe[:EXECute]?

:DIAGnostic:TEST:FRAMe[:EXECute]?

This command executes the specified E5250A test, then returns the test result. For
more info, see Chapter 3.

After :DIAG:TEST:FRAM:EXEC CONT execution, the E5250A status becomes
same as after *RST command execution.

After :DIAG:TEST:FRAM:EXEC FPAN execution, the E5250A status is not
changed.

NOTE

The :DIAG:TEST:FRAM:EXEC FPAN command requires key input from the
front panel. If key is not pressed within 10 seconds, this test result will be fail.

Syntax

:DIAGnostic:TEST:FRAMe[:EXECute]? CONTroller | FPANel

Query response

test_result <newline><^END>

1 : Fail
0 : Pass

Example

OUTPUT @Hp5250;":DIAG:TEST:FRAM? CONT"

Parameter Explanation

CONTroller

Controller test

FPANel

Front panel interface test