To execute self-test using external controller, To execute self-test using external controller -6 – Agilent Technologies Agilent E5250A User Manual

Page 56

Advertising
background image

3-6

Agilent E5250A User’s Guide, Edition 11

Executing Self-Test and Leak Test
Executing Self-Test

To Execute Self-Test using External Controller

You can execute Self-Test from an external controller by using SCPI commands in a
program.

This section describes the requirements, SCPI commands, and a sample program for
Self-Test execution from an external controller. For details about the SCPI
commands, refer to Chapter 7.

Requirements

Controller

GPIB Cable

Relay Test Adapter

BNC Short Caps (3 caps are furnished with each E5255A)

SCPI Commands for Self-Test

Table 3-2 shows the SCPI commands used to execute Self-Test. Keyword of the
SCPI command first level is :DIAG.

The :DIAG commands ignore the E5250A channel configuration mode (NORMAL
or AUTO).

For :DIAG commands that require a card number, you specify 1, 2, 3, 4, or ALL.

If the Fail LED turns on after executing the Self-Test, the E5250A or modules may
be defective. Contact your nearest Agilent Technologies Service Center.

NOTE

Before executing Relay Test, be careful about following:

• Attach the Relay Test Adapter to the E5250A input connectors. Refer to Figure

3-1. Then set the switch on the adapter to the Relay Test position.

• Open all outputs of the plug-in cards as follows: If cable is connected to output

connector of plug-in card, open output at DUT end of cable. If end of cable is
Kelvin connection, disconnect the cable or make non-Kelvin connection at end
of cable.

• For each E5255A installed in the E5250A, connect a BNC short cap to each

E5255A BIAS INPUT connector.