Agilent Technologies Agilent E5250A User Manual

Page 321

Advertising
background image

Agilent E5250A User’s Guide, Edition 11

12-3

SCPI Command Summary

Command

Description

:DIAG:TEST:CARD:CLE card_number

Clears relay test result (pass/fail) of specified card.

card_number: 1, 2, 3, 4, or ALL

:DIAG:TEST:CARD[:EXEC]? card_number

Executes relay test, then returns result:
1: fail card exists, 0: pass

card_number: 1, 2, 3, 4, or ALL

:DIAG:TEST:CARD:STAT? card_number

Returns most recent relay test result:
1: fail, 0: pass, -1: not tested.

card_number: 1, 2, 3, 4

:DIAG:TEST:FRAM:CLE unit

Clears specified test result.

unit: CONT (controller test), or FPAN (front panel
interface test)

:DIAG:TEST:FRAM[:EXEC]? unit

Executes specified test, then returns test result:
1: fail, 0: pass.

unit: CONT (controller test), or FPAN (front panel
interface test)

:DIAG:TEST:FRAM:STAT? unit

Returns most recent test result of specified test:
1: fail, 0: pass, -1: not tested.

unit: CONT: Controller test FPAN: Front panel
interface test

[:ROUT]:BIAS:CHAN:DIS:CARD card_number

Bias disables all channels of specified card.

card_number:
For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL

[:ROUT]:BIAS:CHAN:DIS[:LIST] (@channel_list)

Bias disables the specified channels.

channel_list: Channels to bias disable.

[:ROUT]:BIAS:CHAN:DIS[:LIST]? (@channel_list)

Returns whether specified channels are bias
disabled. 1: disabled, 0: enabled.

channel_list: Channels to check.

Advertising