Connecting the e5250a output, To connect the e5252a output, Connecting the e5250a output -20 – Agilent Technologies Agilent E5250A User Manual

Page 90: To connect the e5252a output -20, Connecting the e5250a output” des

Advertising
background image

4-20

Agilent E5250A User’s Guide, Edition 11

Setting up Measurement Environment
Connecting the E5250A Output

Connecting the E5250A Output

This section describes how to connect the E5252A and E5255A output to connector
plate, your test fixture, and so on.

“To Connect the E5252A Output”

“To Connect the E5255A Output”

WARNING

Turn off the E5250A and all instruments connected to the E5250A. And do not
turn them on until the connection described in this section is completed. If you
ignore this warning, you may be exposed to dangerous voltage.

To Connect the E5252A Output

The E5252A output connector is shown in Figure 4-8. The E5250A input signal
(SMU input or AUX input) appears at the E5252A output connector as shown in
Figure 4-8 when input-output switching path is made.

Figure 4-8

E5252A Output Connector and Output Signal

You can easily connect the E5252A output connector to the connector plate or your
test fixture by using cables shown in Table 4-4.

Table 4-4

E5252A Output Cable

B

A

SMU

Input

AUX

B

A

Force or Sense

Signal line

Guard

Ground

Ground

Agilent Model No.

Description

16494A-001

Triaxial cable, 1.5m (for non-Kelvin connection)

16494A-002

Triaxial cable, 3m (for non-Kelvin connection)

16494B-001

Kelvin Triaxial cable, 1.5m (for Kelvin connection)

16494B-002

Kelvin Triaxial cable, 3m (for Kelvin connection)