Typical data, Supplemental data – Agilent Technologies Agilent E5250A User Manual

Page 301

Advertising
background image

Agilent E5250A User’s Guide, Edition 11

10-7

Specifications

Option 501 Multiplexer

Typical Data

Offset Current

*1

< 0.1 pA (Low Leakage I-V Port)

Channel Crosstalk Capacitance < 0.4 pF per channel (Low Leakage I-V Port)

< 5 pF per channel (Bias Port)

Offset Voltage, EMF at 5min

*2

< 80

μV (Low Leakage I-V Port)

Guard Capacitance

< 180 pF

*3

(Low Leakage I-V Port)

< 60 pF (Low Leakage I-V Port, when the port
is open.)

*1

The offset current when 0 V is applied to all I/O channels.

*2

EMF means Electro Motive Force, and these are offset voltages after
applying voltage and waiting 5 minutes.

*3

This is guard capacitance for an I/O channel for following conditions.
Only one E5255A is installed in mainframe, and only one I/O channel
is connected.

Supplemental Data

Relay Contact Life:

> 10

8

times (Dry Switching Mode)

Settling Time:

< 3.5 sec to settle to within 400 fA of final value
after 10 V is applied (Low Leakage I-V Port)