Diagnostic:test:card:state, Diagnostic:test:frame:clear – Agilent Technologies Agilent E5250A User Manual

Page 185

Advertising
background image

Agilent E5250A User’s Guide, Edition 11

7-17

Command Reference

:DIAGnostic:TEST:CARD:STATe?

:DIAGnostic:TEST:CARD:STATe?

This command returns the most recent relay test result for the specified card.

Syntax

:DIAGnostic:TEST:CARD:STATe? card_number

Query response

test_result <newline><^END>

1 : Fail
0 : Pass
-1 : Not Tested

Example

OUTPUT @Hp5250;":DIAG:TEST:CARD:STAT? 1"

ENTER @Hp5250;A

:DIAGnostic:TEST:FRAMe:CLEar

This command clears test result of the specified E5250A test.

Syntax

:DIAGnostic:TEST:FRAMe:CLEar CONTroller | FPANel

Example

OUTPUT @Hp5250;":DIAG:TEST:FRAM:CLE CONT"

Parameter Explanation

card_number

card number: 1, 2, 3, 4

Parameter Explanation

CONTroller

Controller test

FPANel

Front panel interface test