Agilent E5250A User’s Guide, Edition 11

6-21

Programming the E5250A

Programming Examples

150 OUTPUT @Hp5250;":ROUT:COUP:PORT ALL,'3,5'"
160 OUTPUT @Hp5250;":ROUT:COUP:STAT ALL,ON"
170 !
180 OUTPUT @Hp5250;":ROUT:CLOS "&Channel$
190 OUTPUT @Hp5250;"*OPC?"
200 ENTER @Hp5250;Complete
210 !
220 ! Put Measurement Routine here
230 !
240 OUTPUT @Hp5250;":ROUT:OPEN:CARD ALL"
250 OUTPUT @Hp5250;"*OPC?"
260 ENTER @Hp5250;Complete
270 !
280 OUTPUT @Hp5250;"*RST"
290 !
300 END

Line

Number

Description

150

Sets up SMU3&4 and SMU5&6 input ports as Couple Ports.

160

Turns on the Couple Port mode.

180

Closes relays to create channels specified by Channel$.

190 to 200

Waits until the relay close operation is completed.

220

Here, insert measurement routine for your measurement
instrument.

240 to 260

Opens all relays, and waits until open operation is completed.

280

Resets the E5250A.