E5255a channel list, E5255a channel list -19 – Agilent Technologies Agilent E5250A User Manual

Page 113

Advertising
background image

Agilent E5250A User’s Guide, Edition 11

5-19

Controlling the E5250A

Switch Control Basics

E5255A Channel List

This section describes how to define the channel_list specifically for the E5255A.

For the basic syntax of channel_list parameter, see “Channel List and Channel
Configuration” on page 5-6.

Normal Configuration Mode

To specify a channel, you use five digits as shown in Table 5-7.

Table 5-7

Specifying E5255A Channels in Normal Config. Mode

Slot No.

(E5255A

is

installed)

Block No.

of E5255A

One Channel: 5 digits

Card Number

(first 1 digit)

Channel Number: 4 digits

Input Port No.

(next 2 digits)

Output Port No.

(last 2 digits)

1

1 1

IV

Input:

01

BIAS Input: 51

01 to 08

2

IV Input: 02
BIAS Input: 52

3

IV Input: 03
BIAS Input: 53

2

1 2

IV

Input:

01

BIAS Input: 51

01 to 08

2

IV Input: 02
BIAS Input: 52

3

IV Input: 03
BIAS Input: 53

3

1 3

IV

Input:

01

BIAS Input: 51

01 to 08

2

IV Input: 02
BIAS Input: 52

3

IV Input: 03
BIAS Input: 53

4

1 4

IV

Input:

01

BIAS Input: 51

01 to 08

2

IV Input: 02
BIAS Input: 52

3

IV Input: 03
BIAS Input: 53

Advertising