Agilent E5250A User’s Guide, Edition 11

Controlling the E5250A
Switch Control Basics

Card Number:

One digit: 1 to 4. E5250A slot number in which card is
installed.

Channel Number: Four digits: XXYY.

XX: input port number (depends on the block number).

01, 02, or 03 for IV Input Port Number.

51, 52, or 53 for BIAS Input Port Number.

The input port number depends only on the block number,
and does not depend on which the E5250A SMU Input is
connected to the IV Input.

YY: output port number (01 to 08).

Channel Number assignment is shown in Figure 5-6.

Figure 5-6

E5255A Channel Number in Normal Config. Mode

Block 1

Block 3

Block 2

1

8

7

6

5

4

3

2

BIAS Input
(51)

IV Input

Output

0101 0102

0103 0104 0105 0106 0107 0108

5101 5102

5103 5104 5105 5106 5107 5108

1

8

7

6

5

4

3

2

IV Input

Output

0301 0302

0303 0304 0305 0306 0307 0308

5301 5302

5303 5304 5305 5306 5307 5308

1

8

7

6

5

4

3

2

IV Input

Output

0201 0202

0203 0204 0205 0206 0207 0208

5201 5202

5203 5204 5205 5206 5207 5208

BIAS Input
(53)

BIAS Input
(52)

XXXX

: Channel No.