Agilent E5250A User’s Guide, Edition 11

Programming the E5250A
Programming Examples

160 !
170 ! Put Bias Source Control Routine here.
180 !
190 OUTPUT @Hp5250;":ROUT:CLOS "&Channel$
200 OUTPUT @Hp5250;"*OPC?"
210 ENTER @Hp5250;Complete
220 !
230 ! Put Measurement Routine here
240 !
250 OUTPUT @Hp5250;":ROUT:BIAS:STAT ALL,OFF"
260 OUTPUT @Hp5250;":ROUT:OPEN:CARD ALL"
270 OUTPUT @Hp5250;"*OPC?"
280 ENTER @Hp5250;Complete
290 !
300 OUTPUT @Hp5250;"*RST"
310 !
320 END

Line

Number

Description

170

Here, insert your bias source control routine.

190

Disconnects all output ports of block 1 from the input Bias Port,
then connects the output ports to the IV input port.

200 to 210

Waits until the channel connections are completed.

230

Here, insert measurement routine for your measurement
instrument.

250

Turns off the Bias Mode.

260 to 280

Opens all relays, then connects all output ports of block 1 to the
input Bias Port again, and waits until the connections are
completed.

300

Resets the E5250A.