Agilent Technologies Agilent E5250A User Manual

Page 67

Advertising
background image

Agilent E5250A User’s Guide, Edition 11

3-17

Executing Self-Test and Leak Test

Using the Self-Test Utility

4. Select desired IVn softkey (n=1 to 6) for input port you want to test.

If the E5252A is installed in your E5250A, and the 4156 is used for the Relay
Test, then when you press IV1, IV3, or IV5 softkey, the following message and
softkeys are displayed:

Is IVn Port a Kelvin connection?

Where n=1, 3, or 5.

Select the YES or NO softkey.

The following message and the softkeys are displayed:

Select SMU channel for IVn leak measurement by softkey.

Where n=1 to 6.

Softkey Description

YES

Declares that the port specified by the IVn softkey is used
for the Kelvin connection.

NO

Declares that the port specified by the IVn softkey is used
for the non-Kelvin connection.

CANCEL

Cancels the Leak Test execution.

Softkey Description

SMUm
(m= 1 to 6)

Declares that you will connect the 4155/4156 SMUm to the
E5250A port specified by the IVn softkey. SMU5 and
SMU6 are displayed only if 5th and 6th SMU are installed
in the 41501 expander of the 4155/4156.

CANCEL

Cancels the Leak Test execution.