Agilent Technologies Agilent E5250A User Manual

Page 262

Advertising
background image

9-6

Agilent E5250A User’s Guide, Edition 11

Executing Sample Programs
Vth and Capacitance Measurement Program

Display Setup:

Also the DATA VARIAVLES fields set VTH and BETA.

Auto Analysis Setup:

*LINE1: [ TANGENT

] line on [ Y1 ]

at a point where

[ GRAD

] = [ MAX(GRAD)

]

-----------------------------------------------------------

*LINE2: [

]

-----------------------------------------------------------

*MARKER:

At a point where

[ GRAD

] = [ MAX(GRAD)

]

[

]

---------------------

*Interpolate: [OFF]

Capacitance Measurement Setup

Measurement setup of the E5250A and 4284A for this program is as follows:

E5250A/E5252A Setup:

4284A Setup:

Xaxis Y1axis Y2axis

NAME

Vg

SQRTId

GRAD

SCALE

LINEAR

LINEAR

LINEAR

MIN

0 V

0

0

MAX

5 V

200 m

80 m

Slot No

Channels

1

10905 and 11006

2

not used.

3

not used.

4

not used.

Measurement Frequency

100 kHz

Signal Level

0.03 V

Measurement Function

Cp-G