Typical data, Supplemental data – Agilent Technologies Agilent E5250A User Manual

Page 299

Advertising
background image

Agilent E5250A User’s Guide, Edition 11

10-5

Specifications

Option 001 Matrix Switch

Typical Data

Offset Current

*1

< 0.1 pA (Low Leakage I-V Port)

< 1000 pA (General I-V Port)

Channel Crosstalk Capacitance < 0.3 pF per channel

Offset Voltage, EMF at 5 min

*2

< 80

μV (Low Leakage I-V Port)

< 110

μV (General I-V Port)

< 110

μV (C-V Port, HF Port)

Guard Capacitance

*3

< 145 pF (Low Leakage I-V Port)

< 123 pF (General I-V Port)

Additional Capacitance
Measurement Error

*4

< ± 1 % ± 0.5 pF (C-V Port)

*1

The offset current when 0V is applied to all I/O channels.

*2

EMF means Electro Motive Force, and these are offset voltages after
applying voltage and waiting 5 minutes.

*3

This is guard capacitance for an I/O channel for following conditions.
Only one E5252A is installed in mainframe, and only one I/O channel
is connected.

*4

Additional error for the following measurement conditions. Applicable
for the data compensated by the capacitance compensation routine.

Frequency: 1 kHz to 1 MHz

Measurement Range: Maximum 1000 pF

Measurement Terminal: At the end of Agilent 16495A-001 (1.5 m
triaxial cable) connected to the E5252A output terminal.

Supplemental Data

Bandwidth (at -3dB)

< 10 MHz (C-V Port, HF Port)

Relay Contact Life

> 10

8

times (Dry Switching Mode)

Settling Time

< 3.5 sec to settle to within 400 fA of final value
after 10 V is applied (Low Leakage I-V Port)

Advertising