Agilent E5250A User’s Guide, Edition 11

3-11

Executing Self-Test and Leak Test

Using the Self-Test Utility

To Start the Self-Test Utility

1. Before turning on the 4155/4156 and E5250A, connect the GPIB cable between

the E5250A and the 4155/4156.

2. Turn on the 4155/4156 and E5250A.

3. Set the 4155/4156 to "SYSTEM CONTROLLER" mode on SYSTEM:

MISCELLANEOUS page.

4. Insert the E5250A Program Disk into the 4155/4156 disk drive.

5. Press the

Display

key twice to display the "All IBASIC" screen. This key is

located in the IBASIC area on the 4155/4156 front panel.

6. Select GET softkey on the 4155/4156 screen. The following is displayed:

GET "_"

7. Type SELFTEST between the double quotations of GET statement. Then press

Enter

key. Wait until the 4155/4156 finishes getting the file.

8. Press

Run

key, which is in the IBASIC area on the 4155/4156 front panel. The

following message is displayed:

Input Agilent E5250A GPIB Address (default 722):?

9. To use default GPIB address 722, just press the

Enter

key. If not, type the GPIB

address of your E5250A, then press

Enter

key. MAIN MENU of the E5250A

Self-Test Utility and following softkeys are displayed on the 4155/4156:

Softkey Description

SELF TEST

Displays the SELF TEST MENU.

LEAK TEST

Displays the LEAK TEST MENU.

QUIT

Quits the Self-Test Utility.