Agilent Technologies Agilent E5250A User Manual

Page 94

Advertising
background image

4-24

Agilent E5250A User’s Guide, Edition 11

Setting up Measurement Environment
Measurement Cable Length

Table 4-6

Guard Capacitances of E5250A Measurement Environment (Typical)

Connection

Agilent Model/Part No.

Cable

Length

Guard

Capacitance

from SMU to
E5250A

16494A-003 (for non-Kelvin)

80 cm

75 pF

16494B-003 (for Kelvin)

80 cm

90 pF

16494C-001 (for 4142B)

1.5 m

140 pF

E5250A and
Plug-in Card

E5252A (only 1 card is installed)

145 pF

E5252A (more 1 card)

8 pF/extra card

E5255A (only 1 card is installed)

180 pF

E5255A(more 1 card)

60 pF/extra card

from Card
Output to
Connector
Plate

16494A-001 (for non-Kelvin)

1.5 m

125 pF

16494A-002 (for non-Kelvin)

3 m

240 pF

16494B-001 (for Kelvin)

1.5 m

140 pF

16494B-002 (for Kelvin)

3 m

260 pF

16494C-001 (for 4142B)

1.5 m

140 pF

16494C-002 (for 4142B)

3 m

260 pF

16494D-001

1.5 m

125 pF

16494D-002

3 m

240 pF

16494E-003

3 m

240 pF

from
Connector
Plate to DUT

Agilent Part No. 8121-1191

X m

130 pF per m