2-10

Agilent E5250A User’s Guide, Edition 11

Installation
Installing the E5250A

To Configure E5255A

The Agilent E5255A has three 2-input 8-output multiplexers. You can configure
several multiplexers by modifying some internal connections. Component locations
on the E5255A are shown in Figure 2-2, and the relationship of the multiplexer
blocks and the components is shown in Table 2-1. Table 2-2 shows the initial setting
of the E5255A internal connections. The setting does not depend on the slot in
which the module is installed.

Table 2-1

Multiplexer Blocks and Components

Table 2-2

Agilent E5255A Initial Setup

To Modify IV Input

IV input connection is defined by the connection of IVin cable and the setting of
DIP switch shown in Figure 2-2. For actual examples of modifications, see
“E5255A Configuration Examples”.

Each DIP switch has four bit switches. Bits 1 to 3 are used to specify which SMU
input connector is connected to the IV input. Bit 4 of DIP switch is used to specify
whether the protective resistors are 0

Ω or not. The DIP switches must be set

properly after you complete the internal connection. Make sure you set the DIP
switches correctly. If not, dangerous voltages may be applied to unexpected output
connectors.

Multiplexer

block

Resistor

holder

BIASin

BIASout

IVin

cable

IVout

connector

DIP SW

Block 1

holder 1

-

BIASout1

IVin1

IVout1

DIP SW1

Block 2

holder 2

BIASin2

BIASout2

IVin2

IVout2

DIP SW2

Block 3

holder 3

BIASin3

-

IVin3

IVout3

DIP SW3

Block number

IVin cable connection

BIASin/BIASout

Protective resistor

1

SMU1

BIAS INPUT 51

0

Ω

2

SMU2

BIAS INPUT 52

0

Ω

3

SMU3

BIAS INPUT 53

0

Ω