Agilent E5250A User’s Guide, Edition 11

11-11

Error Messages

E5250A Specific Error Messages

E5250A Channel Related Errors

Error

Number

Error Message and Description

2000

Invalid card number

Wrong card number is specified in card number or channel list
parameter. Make sure card is properly installed in the correct
E5250A slot.

2001

Invalid channel number

Wrong channel number is specified in the channel list. Confirm the
channel number, the card configuration, and the configuration
mode of the E5250A.

2006

Command not supported on this card

Command was used that is not supported by the card.

2009

Too many channels in channel list

Too many channels are specified in the channel list. Maximum
number of channels:

E5252A: number of cards

× 120

E5255A: number of cards

× 48

2011

Empty channel list

No channel list is specified.

2012

Invalid channel range

Wrong channel list is defined. Confirm the syntax of the channel
list, the card configuration, and the configuration mode of the
E5250A.