To make interlock circuit, To make interlock circuit -8 – Agilent Technologies Agilent E5250A User Manual

Page 78

Advertising
background image

4-8

Agilent E5250A User’s Guide, Edition 11

Setting up Measurement Environment
Mounting Connectors Directly

To Make Interlock Circuit

The interlock circuit is to prevent electric shock when touching measurement
terminals.

You must install an interlock circuit on shielding box to prevent dangerous voltages
when door of the shielding box is open.

Figure 4-2 shows the pin assignments of the interlock connector mounted on your
own connector plate or test fixture.

Figure 4-2

Interlock Connector Pin Assignments

WARNING

Dangerous voltages of up to the maximum voltage of SMUs may be present at
force, guard, and sense terminals when the interlock terminals are shorted.

To Install Interlock Circuit

Install the interlock circuit as follows.

1. Mount two mechanical switches on your shielding box, so that the switches

close when the door of the shielding box is closed, and open when the door is
opened. For the dimensions of the switch, refer to Figure 4-3 and Figure 4-4.

2. Use wire to connect the two switches in series between pin number 1 and 2 (or

3) of the interlock connector. See Figure 4-2.

For example, Agilent 4155/4156 is connected to the interlock circuit, it cannot force
more than ±40 V when the door is open. When door is closed, it can force more than
±40 V.