Agilent E5250A User’s Guide, Edition 11

4-13

Setting up Measurement Environment

Mounting Connectors Directly

To Connect Connectors to DUT

Following figure shows a connection example from the connector to DUT. Use
low-noise coaxial cable (Agilent part number: 8121-1191).

CAUTION

Never connect the guard terminal to any output, including circuit common, chassis
ground, or the guard terminal of any other unit. Doing so may result in an
emergency condition.

Kelvin connections

non-Kelvin connections

This connection is available only for the
Kelvin connectors. This connection can cancel
effects of cable resistance by connect sense
line and force line as close as possible to DUT
terminal.

Following connection is for the Kelvin
connectors. For the triaxial connectors or the
8 ch shielded connectors, ignore SENSE
terminal, and make connection only for
FORCE terminal. Measurement results
include residual resistance from the
connection cable.

to
DUT

Coaxial cable

Triaxial
connector

Insulator

Plate

Guard

Sense

Common

Guard

Force

Common

to
DUT

Coaxial cable

Triaxial
connector

Insulator

Plate

Guard

Sense

Common

Wire

Guard

Force

Common