5-14

Agilent E5250A User’s Guide, Edition 11

Controlling the E5250A
Switch Control Basics

The two output ports are determined by the channel_list parameter of the
:ROUT:CLOS command, and the lower output port number is always odd.

So, when you execute :ROUT:CLOS or :ROUT:OPEN command for the couple
port, four relays operate synchronously as shown in Figure 5-3.

For example, assume that input ports 1 and 2 are coupled, and that the Couple Port
mode is ON:

1. If :ROUT:CLOS command specifies to connect channel 0105, relays are closed

to realize two channels (0105 and 0206).

2. If :ROUT:OPEN command specifies to disconnect channel 0105, all four relays

between the two input ports and two output ports are opened.

Figure 5-3

Couple Port Relay Connections

NOTE

For each E5252A installed in the E5250A, you can independently specify the same
or different couple ports.

INPUT

OUTPUT

INPUT

OUTPUT

2XM-1

2XM

2XN

2XN-1

2XM-1

2XM

2XN

2XN-1

1. Connect condition

2. Disconnect condition

Close
Open

N=1 to 5

M=1 to 6 for Normal Config.

M=1 to 24 (max) for Auto Config.