9-32

Agilent E5250A User’s Guide, Edition 11

Executing Sample Programs
HCI Measurement Program

To Execute Analysis Program

Before running the program, do the following:

1. Insert the E5250A Program Disk into the computer's disk drive.

2. Specify the mass storage device (disk drive) the program disk is inserted by

using MSI statement. For example, if the device is ":CS80,702,1", enter the
following command:

MSI ":CS80,702,1"

3. Get the SP_MUX_A program from the program disk.

GET "SP_MUX_A"

4. Eject the program disk.

5. Insert the diskette which saves test result data into the computer's disk drive.

6. Specify the directory which stores the test result data files. For example, if the

directory is /test_data, enter

MSI "/test_data"

.

Execute the SP_MUX_A program as follows:

1. Press the RUN function key on the computer. The following message is

displayed.

"Enter number of devices (default:24)"

2. Enter the number of devices to be analyzed. Default number is 24.

Then the following softkeys are displayed.

Tdc is the "time to target". The SP_MUX_A program extracts Tdc from the
measurement data that was saved by SP_MUX program. Tdc is the time when
the device parameter reached its termination condition.

Softkey description

Idlin

Extract Tdc from the Idlin data for all valid devices

Gmmax

Extract Tdc from the Gmmax data for all valid devices

Vtext

Extract Tdc from the Vtext data for all valid devices

Vtci

Extract Tdc from the Vtci data for all valid devices

All

Extract Tdc for all these parameters for all valid devices

End

Quit the program