Agilent Technologies Agilent E5250A User Manual

Page 266

Advertising
background image

9-10

Agilent E5250A User’s Guide, Edition 11

Executing Sample Programs
Vth and Capacitance Measurement Program

Execute the program as follows:

1. LOAD the Vth and C measurement program as following example.

LOAD "P_MAT"

2. Link the Capacitance Compensation Routine by using the LOADSUB ALL

statement as in following example.

LOADSUB ALL FROM "C_COM"

3. Eject the E5250A Program Disk from the disk drive connected to the computer.

4. Insert the E5250A Program Disk into the 4155/4156's disk drive. Setup file

E5252.MES is stored on the disk and is used to set up 4155/4156 for the Vth
measurement.

5. RUN the program.

6. After the measurement, measurement results are displayed on the computer as

shown in example in Figure 9-2.

Figure 9-2

Example of Vth and C Measurement Results

Comment

: Vth & C Meas. Sample Program

Test Time : 30 Sep 1995 10:29:27

# P/F

Min.

Meas.

Max.

Comment

1

P 10.000E-002 18.255E-001 20.000E-001 MOSFET Vth Measurement

2

F 8.000E-12

2.164E-11

2.000E-11

Capacitance Measurement