Agilent E5250A User’s Guide, Edition 11

6-17

Programming the E5250A

Programming Examples

Program List:

10 ! Example: Using Bias Mode
20 !
30 INTEGER

Complete

40 DIM

Channel$[100]

50 Channel$="(@10101:10108)"
60 !
70 ! Put Instrument Initialization Routine here.
80 !
90 ASSIGN @Hp5250 TO 722
100 OUTPUT @Hp5250;"*RST"
110 OUTPUT @Hp5250;":ROUT:FUNC NCON"
120 OUTPUT @Hp5250;":ROUT:CONN:RULE ALL,FREE"
130 !OUTPUT @Hp5250;":ROUT:BIAS:PORT 1,7"
140 OUTPUT @Hp5250;":ROUT:BIAS:CHAN:ENAB:CARD 1"
150 OUTPUT @Hp5250;":ROUT:BIAS:STAT 1,ON"

Line

Number

Description

40

Declares data size of Channel$, which is used as channel_list
parameter.

50

Defines the desired channel_list.

70

Here, insert the initialization routine for your measurement
instrument.

100

Resets the E5250A.

110

Sets the E5250A to Normal channel configuration mode.

120

Sets the connection rule to FREE.

130

This line is not used (commented) for the E5255A. If you use the
E5252A instead of the E5255A, use this line, which defines AUX
INPUT HF1 to be the input Bias Port.

140

Bias enables all output ports of card 1 for the Bias Mode.

150

Turns on the Bias Mode. All output ports of card 1 are connected to
the input Bias Port.